This page requires that you know the email address that you've been signed up with. If you wish to log a support call please using the link on the right below and then sign in with your site's username and password. If you are unsure about any of this information please contact the Training Academy on 0870 0666960

http://academy.sysmex.co.uk/

  • GROWING YOUR KNOWLEDGE
    GROWING YOUR KNOWLEDGE
  • DEVELOPING YOUR SKILLS
    DEVELOPING YOUR SKILLS
  • HARVESTING EXPERTISE
    HARVESTING EXPERTISE

WBC/BASO channel on XE/XT-Series

The WBC/BASO channel records forward scattered light and side scattered light signals.

  • Forward scattered light, or FSC, provides information about the cell size
  • Side scattered light, or SSC, describes the internal cell structure.

An acidic reagent (stromatolyser-FB) lyses the red blood cells and shrinks white blood cells to bare nuclei, with the exception of basophils, which are not affected. The resulting differences between basophils and other cells are analysed using forward and side scatter information.

The WBC/BASO channel is not available on XS-Series analysers.

The WBC count in CBC mode on XS-Series is performed with the Stromatolyser-4 DL reagent. All red blood cells are lysed by this reagent in the WBC channel. The forward scatter signal is used to count the WBC. The results are displayed in a histogram. In the CBC+DIFF mode, the WBC count is taken from the fluorescence signals from the DIFF channel after treating the cells with Stromatolyser-4 DS (fluoromarker).