This page requires that you know the email address that you've been signed up with. If you wish to log a support call please using the link on the right below and then sign in with your site's username and password. If you are unsure about any of this information please contact the Training Academy on 0870 0666960

http://academy.sysmex.co.uk/

  • GROWING YOUR KNOWLEDGE
    GROWING YOUR KNOWLEDGE
  • DEVELOPING YOUR SKILLS
    DEVELOPING YOUR SKILLS
  • HARVESTING EXPERTISE
    HARVESTING EXPERTISE

Flocculated cryoglobulin in phase contrast microscopy image

Flocculated cryoglobulin in phase contrast microscopy image

In the impedance channel of a haematology analyser a falsely elevated platelet concentration of 850,000/μL was measured. This was caused by precipitated cryoglobulin in cryoglobulinaemia. The flocculated cryoglobulin is visible in the phase contrast microscope in between the red blood cell 'rouleaux'.