CS-5100
The CS-5100 is the latest, fully automated coagulation analyser produced by Sysmex. It has been specifically designed for high throughput laboratories. The CS-5100 provides improved quality of results and will improve efficiency within your laboratory.
The CS-5100 supports both closed and open vial sampling and simultaneously measures multiple parameters using the multi-wavelength detection system. It provides automatic reflex, repeat and re-analysis and re-dilution of samples and is compatible with automated sample tracking, eliminating the need for robotic arms; ensuring that the CS-5100 is a reliable and highly cost effective solution.
- Pre-analytical sample checks-Increased result security, reduced effect of interfering substances
- HIL check for interfering substances-provides objective detection of haemolysis, Icterus and Lipeamia for ultimate sample integrity
- Advanced Reagent Management- Multiple vials of the same reagent with Auto QC on vial change
- Systemisation- Compatible with laboratory automated transport systems (LAS)
- Automated vWF:RCof Assay-Significant precision and accuracy improvements to the gold standard platelet aggregation assay
- Calibration and Assay Performance- Automatic recognition of the appropriate calibration curve
- Primary Sample volume check- automatic detection of under and over filled samples
- Automatic wavelength switch in the presence of significant interference
- User definable flagging of compromised samples
- Information Technology-Integrated result management with Sysmex information systems
- Eight platelet aggregation detectors with integrated stirring
- Principles Transmitted light at multiple wavelengths: 340, 405, 575, 660, and 800 nm for clotting, immunoassay and aggregation.
- Detectors 20 channel for clotting, chromogenic and immunoassays, 8 channels for platelet aggregation. Transmitted light readings taken every o.1s wavelength.
- Throughput PT 400 - Tests/hr PT/APTT - 357 Tests/hr PT/APPTT/Fbg - 360 Tests/hr PT/APTT/Fbg/D-Dimer - 268 Tests/hr
- Quality Control Levy Jennings Control Multi-rule (Westguard Rule) monitoring 1200 data points x 750 parameters in memory
- Data Storage 10,000 sample results with reaction curves
- Power Requirements 200-240V (50/60Hz)
- Max Power Consumption Main Unit - 1700 VA max Pneumatic - 250 VA max
- Dimensions Main Unit Approx - 1030W x 1150D x 1350H (mm) 267kg PC/IPU and Cabinet - 500W x 890D x 920H (mm) 67kg