HISCL™ ADAMTS13 Activity Assay Kit
ADAMTS13 Activity testing now available on your CN-3500/CN-6500
-
Rapid results – ADAMTS13 activity results in approximately 20 minutes, supporting timely clinical decision-making when TTP is suspected.
-
Testing consolidation – expand your specialist haemostasis testing portfolio using chemiluminescence technology on a single routine coagulation platform.
-
Long onboard stability – reagents remain stable onboard for up to 60 days, helping laboratories manage specialist testing efficiently.
-
Confidence at critical decision points – excellent clinical sensitivity and specificity for TTP, supported by quality control levels positioned around clinically important decision points.
Bringing together speed, automation and workflow efficiency, the HISCLTM ADAMTS13 Activity Assay provides laboratories with a new approach to this critical specialist haemostasis test.
ADAMTS13 Activity testing with HISCL CLEIA technology
The measurement of ADAMTS13 Activity, alongside clinical pre-probability scores and clinical presentation, is crucial in the diagnosis of Thrombotic Thrombocytopenic Purpura (TTP). The new HISCL ADMATS13 Activity Assay Kit provides a highly sensitive and specific CLEIA assay to aid the diagnosis of TTP utilising the internationally recognised 10 IU/dL cut-off1.
Thrombotic Thrombocytopenic Purpura (TTP)
ADAMTS13 is an enzyme that regulates blood clot formation by cleaving ultra-large von Willebrand factor (vWF) multimers, preventing excessive platelet aggregation. Severe ADAMTS13 deficiency (congenital or immune mediated) can lead to thrombotic thrombocytopenic purpura (TTP), a rare, life-threatening blood disorder characterised by platelet-rich microvascular thrombi, thrombocytopenia and haemolytic anaemia. Early diagnosis and prompt treatment are critical, as untreated TTP has a mortality rate exceeding 90%, whereas modern therapies have significantly improved patient survival.
Reference:
Kubo M, Konko K, Kinoshita E, Uemae S, Kobayashi K, Hayashi Y, Kan A, Fujimura Y, and Matsumoto M. A novel automated chemiluminescent enzyme immunoassay for ADAMTS-13 activity enables accompanying measurements of the inhibitory autoantibodies. J Thromb Haemost. 2025 Mar; 23(3): 957–967.

HISCL ADAMT13 Assay Principle – powered by CLEIA
This kit quantifies ADAMTS13 activity in the sample based on a 2-step CLEIA principle.
- The biotinylated-vWF73 peptide in the R1 reagent is cleaved by the ADAMTS13 in the sample. The cleaved portion of the biotinylated-vWF73 peptide is captured by the streptavidin-coated magnetic particles (MP) in the R2 reagent.
- After B/F (Bound and Free) separation, alkaline phosphatase (ALP)-conjugated anti-cleaved vWF73 peptide antibody (anti-N10) in the R3 reagent bind to the cleaved biotinylated-vWF73 peptide on the MP.
- After B/F separation, the ALP-conjugated antibody on the MP reacts with CDP-Star™ substrate in the R5 reagent to produces a luminescent signal.
- The light emission is directly proportional to the ADAMTS13 activity.
The signal from an unknown sample is compared to a calibration curve, created using a sample containing a known and traceable amount of ADAMTS13 activity (HISCL ADAMTS13 Calibrator) to generate a result in international units per decilitre (IU/dL).
Measurement item | ADAMTS13 Activity |
Packaging | 50 tests/package |
Measurement principle | Chemiluminescent enzyme immunoassay (CLEIA) |
Purpose of use | An in vitro diagnostic reagent kit for the quantification of ADAMTS13 activity in human plasma |
Sample type | Human citrated plasma |
Measurement range | 0.50 – 120.0 IU/Dl |
Cut-off value | 10 IU/dL |
Time to result | 20 minutes |
| On-board stability | 60 days |
Applicable instruments | Sysmex Automated Blood Coagulation Analyser CN-3500/CN-6500 |